Table 2.
Accuracy of Estimators TE, RE, and A in Phantoms, Together with Corresponding Limiting Values for Each Couple of Image Type and Registration Algorithm
| Phantom | Registration | Algorithm | TE | TEmax | RE | REmin | A | Amax |
|---|---|---|---|---|---|---|---|---|
| Alderson Rando | CT-SPECT | MIa | 0.36 | 0.55 | 0.9981 | 0.9969 | 0.0040 | 0.0119 |
| CT-MRI | LCb | 0.06 | 0.34 | 0.9999 | 0.9988 | 0.00003 | 0.0029 | |
| CT-MRI | MI | 0.10 | 0.9999 | 0.0001 | ||||
| Hoffman | CT-SPECT | MI | 0.43 | 0.55 | 0.9991 | 0.9969 | 0.0039 | 0.0119 |
| CT-MRI | LC | 0.22 | 0.34 | 0.9998 | 0.9988 | 0.0005 | 0.0029 | |
| CT-MRI | MI | 0.21 | 0.9992 | 0.0009 |
aMI: Mutual information
bLC: Local correlation