Accuracy of Estimators TE, RE, and A in Phantoms, Together with Corresponding Limiting Values for Each Couple of Image Type and Registration Algorithm

Phantom Registration Algorithm TE TEmax RE REmin A Amax
Alderson Rando CT-SPECT MIa 0.36 0.55 0.9981 0.9969 0.0040 0.0119
CT-MRI LCb 0.06 0.34 0.9999 0.9988 0.00003 0.0029
CT-MRI MI 0.10 0.9999 0.0001
Hoffman CT-SPECT MI 0.43 0.55 0.9991 0.9969 0.0039 0.0119
CT-MRI LC 0.22 0.34 0.9998 0.9988 0.0005 0.0029
CT-MRI MI 0.21 0.9992 0.0009

aMI: Mutual information

bLC: Local correlation