Table 1.

Values of retention time, regression equations, correlation coefficients for linear calibration plots, and limits of detection obtained for examined PFCAs in reversed-phase HPLC determination with fluorescence detection employing derivatization with 3-BrAC (without additional preconcentration)

PFCA Retention time (min) Regression equations Correlation coefficient LOD (μg·l−1)
C2 6.5 y = 336356 x 0.9997 7.5
C3 9.1 y = 356820 x 0.9998 5.5
C4 11.3 y = 640191 x 0.9997 6.0
C5 13.3 y = 224169 x 0.9996 6.8
C6 15.0 y = 354849 x 0.9995 6.5
C7 16.5 y = 222639 x 0.9997 6.6
C8 17.9 y = 190238 x 0.9998 4.6
C9 19.1 y = 251250 x 0.9998 4.3
C10 20.2 y = 111682 x 0.9997 6.8
C11 21.1 y = 88769 x 0.9995 6.5
C12 22.0 y = 95421 x 0.9998 7.5