Published in final edited form as: Biomed Chromatogr. 2009 Dec;23(12):1251–1258. doi: 10.1002/bmc.1247

Table 1.

Intra-day accuracy and intra-day precision of imatinib quantification in human blood, leukemia cells and cell culture media

Human whole
blood
Human leukemia
cells
Cell culture
media
Intra-day accuracy LLOQ 100.9% 98.1% 93.2%
0.1 ng/mL 108.6% 100.1% 100.1%
1 ng/mL 94.7% 100.0% 100.2%
10 ng/mL 107.8% 107.6% 98.8%
ULOQ 92.8% 86.5% 94.24%
Intra-day precision LLOQ 8.8% 13.2% 14.9%
0.1 ng/mL 7.2% 6.8% 6.8%
1 ng/mL 6.7% 3.1% 7.3%
10 ng/mL 8.2% 7.0% 2.4%
ULOQ 4.3% 5.4% 11.0%
In-process stability LLOQ 114.1% 118.1% 107.4%
0.1 ng/mL 97.3% 114.0% 113.3%
1 ng/mL 93.1% 111.0% 113.7%
10 ng/mL 102.0% 109.7% 115.9%
ULOQ 99.6% 112.0% 105.4%

LLOQ, lower limit of quantitation; ULOQ, upper limit of quantitation. LLOQ for human whole blood and leukemia cells: 0.03 ng/mL and for cell culture media 0.01 ng/mL; in-process stability measured after 24 h at +4°C; ULOQ, 75 ng/mL.