. 2011 Jan 1;5(1):99–106. doi: 10.1177/193229681100500114

Table 2.

Performance Characteristics of the FreeStyle Navigator CGM with the TRUstart Algorithma

Point EGA
Zone A 80.3 (6306)
Zone B 18.0 (1410)
Zone C 0.2 (16)
Zone D 1.5 (120)
Zone E 0 (0)
Rate EGA
Zone A 79.1 (6212)
Zone B 16.1 (1266)
Zone C 2.1 (162)
Zone D 2.0 (158)
Zone E 0.7 (54)
G-EGA (combined grids to give overall CGM clinical accuracy)
Hypoglycemia (≤70 mg/dl) Euglycemia (71–180 mg/dl) Hyperglycemia (>180 mg/dl) All data
Clinically accurate 53.6 (162) 95.6 (4393) 94.7 (2800) 93.7 (7355)
Benign errors 7.0 (21) 3.7 (168) 3.1 (91) 3.6 (280)
Clinical errors 39.4 (119) 0.7 (32) 2.2 (66) 2.8 (217)
Statistical measures of concordance of CGM glucose with reference glucose values
Mean ARD 21.6 (302) 15.5 (4593) 10.5 (2957) 14.5 (7852)
Median ARD 17.5 (302) 11.8 (4593) 8.3 (2957) 10.7 (7852)
ISO accuracy 44.0 (133) 72.3 (3319) 87.4 (2584) 76.9 (6036)
a

Table entries are % (n). Zone A represents clinically accurate; zone B represents benign or no treatment errors; zone C represents overcorrection errors; zone D represents a potentially dangerous failure to treat; and zone E represents erroneous treatment.