Table 2.

CONTOUR TS Results within ISO 15197 Performance Criteria

Blood sample Meter operator Percentage of results within
±20% or ±15 mg/dl of YSI result ±15% or ±10 mg/dl of YSI result
Capillary Subject 97.9% (411/420) 91.7% (385/420)
HCP 98.6% (412/418) 93.3% (390/418)
All results 98.2% (823/838) 92.5% (775/838)
Venous HCP 99.8% (623/624) 97.9% (611/624)