Table 2.

Effect of free-fall test and static load test on safety cap removal force

Sample number (n) Mean ± SD cap removal force (N) P-value
Without preconditioning – standard conditions (control)
Syringe AAI – protection cap 20 16.58 ± 5.86
Syringe AAI – safety cap 15 20.23 ± 6.61
Cartridge AAI – safety cap 15 9.56 ± 2.36 P < 0.001 versus syringe AAI – safety cap
Preconditioning – free-fall test
Syringe AAI – safety cap 15 21.07 ± 1.19 P = 0.690 versus control
Cartridge AAI – safety cap 15 9.98 ± 0.68 P = 0.653 versus control
Preconditioning – static load test
Syringe AAI – safety cap 15 18.89 ± 2.203 P = 0.633 versus control
Cartridge AAI – safety cap 15 38.39 ± 1.458 P < 0.001 versus control