Table 1.
Lowest stimulation thresholds for model PTNs (P2, P3, P4), cortical interneurons (t1, n2 and n3) and association fibers (a1, a2) as a function of waveform type. The table also shows the sites and mechanisms of activation and the phase of the waveform at which stimulation occurred.
Model Neuron (diameter) | Pulse type | Lowest threshold | Stimulation | Phase of waveformb | |||
---|---|---|---|---|---|---|---|
A/μs | % of MSOa | Site | Mechanism | ||||
P2 (20 μm) | Mono | PA | 97.7 | 57% | Axonal bend in white matter | Charge accumulation at the bend | 1st 2nd 2nd 3rd |
AP | 263.1 | 154% | |||||
Bi | AP-PA | 73.4 | 72% | ||||
PA-AP | 107.3 | 106% | |||||
P3 (20 μm) | Mono | PA | 90.9 | 53% | |||
AP | 252.9 | 148% | |||||
Bi | AP-PA | 70.7 | 70% | ||||
PA-AP | 98.4 | 97% | |||||
P4 (20 μm) | Mono | PA | 105.9 | 62% | |||
AP | 291.4 | 170% | |||||
Bi | AP-PA | 81.5 | 80% | ||||
PA-AP | 114.4 | 113% | |||||
Collateral of P1 (6 μm collateral, 20 μm main axon) | Mono | PA | 64.8 | 38% | Collateral termination | Charge accumulation at axonal termination | |
AP | 165.5 | 97% | |||||
Bi | AP-PA | 48.2 | 47% | ||||
PA-AP | 72.2 | 71% | |||||
t1 (6μm) | Mono | PA | 65.7 | 38% | Axonal termination (gray matter) | Charge accumulation at axonal termination | |
AP | 162.6 | 95% | |||||
Bi | AP-PA | 48.0 | 47% | ||||
PA-AP | 72.7 | 72% | |||||
n2 (6μm) | Mono | PA | 116.7 | 68% | |||
AP | 289.8 | 170% | |||||
Bi | AP-PA | 85.1 | 84% | ||||
PA-AP | 127.5 | 125% | |||||
n3 (6 μm) | Mono | PA | 127.9 | 75% | |||
AP | 318.1 | 186% | |||||
Bi | AP-PA | 93.4 | 92% | ||||
PA-AP | 139.7 | 137% | |||||
a1 (10 μm) | Mono | PA | 171.7 | 101% | Axonal terminationc (gray matter, M1) | Charge accumulation at axonal | 2nd 1st 3rd 2nd |
AP | 104.8 | 61% | |||||
Bi | AP-PA | 106.7 | 105% | ||||
PA-AP | 79.1 | 78% | |||||
a2 (20 μm) | Mono | PA | 70.2 | 41% | |||
AP | 40.6 | 24% | |||||
Bi | AP-PA | 41.6 | 41% | ||||
PA-AP | 31.1 | 31% |
All values above 100 % for the Magstim 200 stimulator, and 120 % for the Magstim Rapid stimulator are outside the range of values that the stimulator can provide.
Phase of pulse waveform for which stimulation occurred. Each row corresponds to the pulse types displayed in the second column.
For all pulse types, except monophasic PA, the threshold for activation at the bend after the fiber leaves the somato-sensory cortex was very similar to this one.