Published in final edited form as: J Arthroplasty. 2010 Sep 18;26(5):796–801. doi: 10.1016/j.arth.2010.07.016

Table 1.

Case data and inspection results

Case #Cr TIV(y) G Age (y) W (kg) Revision Diag. AnA (°) AbA (°) Dia (mm) Rev. Hist.
A 4 0.08 F 74 68 Infection 9 52.6 36 Primary
B 2 0.07 M 60 101 Instability 11 41.4 36 Primary
C 1 4.11 F 69 82 Infection 14 46 28 Primary
D 1 2.09 F 52 63 Subluxation/Acet. Loose 31 48.6 36 2nd Rev.
E 2 0.01 M 51 125 Infection 16 49.8 40 1st Rev.
F 5 1.00 F 72 82 Pain 9 45.2 36 Primary
G 0 6.26 F 46 79 Fem. Loose 20 41.6 28 1st Rev.
H 0 0.80 F 62 85 Dislocation 11 46.6 32 1st Rev.
I 0 0.03 F 83 110 Dislocation 33 55.2 32 Primary

Num. cracks (#Cr). Time In Vivo (TIV), Anteversion (AnA), Abduction (AbA)

All implants are Zimmer Trilogy™ with Longevity™ highly cross-linked UHMWPE with elevated rim