. 2011 Feb 11;38(3):1196–1206. doi: 10.1118/1.3544658

Table 1.

Comparison of CTDI100 (mGy∕100 mA s) obtained from measurements and simulations for head and body CTDI phantoms and 10 mm collimation for the tube potentials of 80, 100, and 120 kVp.

Collimation width (mm) Phantoms Measurement position in CTDI phantoms Measured CTDI100 (mGy∕100 mA s) Simulated CTDI100 (mGy∕100 mA s) Percent error (%)
80 kVp Head phantom Center 5.3 5.1 −4.1
12 o’clock 6.9 6.6 −3.3
3 o’clock 6.5 6.3 −3.4
6 o’clock 5.7 5.7 −0.3
Body phantom Center 1.0 0.9 −5.7
12 o’clock 2.7 2.5 −6.2
3 o’clock 2.4 2.4 1.1
6 o’clock 2.1 2.1 0.4
100 kVp Head phantom Center 9.7 9.2 −5.2
12 o’clock 11.8 11.4 −3.7
3 o’clock 11.3 10.8 −4.3
6 o’clock 9.9 9.9 −0.3
Body phantom Center 2.1 2.0 −6.5
12 o’clock 5.0 4.6 −7.2
3 o’clock 4.6 4.6 0.1
6 o’clock 4.0 4.0 0.0
120 kVp Head phantom Center 13.8 13.1 −5.2
12 o’clock 16.8 15.6 −7.2
3 o’clock 15.6 14.9 −4.6
6 o’clock 13.6 13.6 0.1
Body phantom Center 3.4 3.2 −7.9
12 o’clock 7.4 6.7 −8.6
3 o’clock 6.8 6.7 −2.0
6 o’clock 6.0 5.9 −1.0