Table 2.

ESAK with BS, Tube Potential (Peak Kilovoltage), and Added Filtration (Millimeter Cu), SNR, and HCSR Parameter for Cine Mode and All PMMA Thicknesses Used in the FD X-ray System (Pediatric Patients Protocol)

PMMA (cm) Acquisition mode ESAK (µGy/fr) Tube potential (kVp) Filter (mm Cu) SNR for matrix 512 (for circle 2) SNR for matrix 1,024 (for circle 2) HCSR for matrix 512 (for 8th group of the central grid) HCSR for matrix 1,024 (for 8th group of the central grid)
4 CI 2.1 62 0.9 8.9 ± 0.1 7.0 ± 0.2 6.9 ± 0.1 128.0 ± 1.4
8 CI 5.8 63 0.6 8.2 ± 0.9 6.7 ± 0.2 6.6 ± 0.1 118.5 ± 3.0
12 CI 24.9 67 0.3 6.9 ± 0.2 5.5 ± 0.4 5.7 ± 0.1 102.2 ± 1.4
16 CI 44.8 68 0.2 6.6 ± 0.1 5.4 ± 0.2 5.3 ± 0.2 95.7 ± 1.2

FOV 25 cm