Table 3.
Intra- and inter- day accuracy and precision of QC samples at the lowest concentration for 5 analytes in 6 media (n=5 each). Accuracy is determined as the percentage deviation of calculated concentrations from the nominal concentrations; Precision is determined as the coefficient of variation (CV) of the 5 measurements. Additional data using higher concentrations can be seen in Supplementary Table 3.
A. Intraday variability | |||||||
---|---|---|---|---|---|---|---|
Matrix | Mobile phase B | Cell incubation medium | Cell lysate, LGLAa | Cell lysate, HGHAb | Rat plasma | Rat urine | |
ARG | Concentration (μM) | 0.2 | 0.2 | 8 | 10 | 8 | 8 |
Accuracy (%) | −8.30 | −1.80 | 8.80 | 6.20 | −0.33 | 12.4 | |
Precision (%) | 3.42 | 11.1 | 3.62 | 4.18 | 2.59 | 1.77 | |
| |||||||
15N4-ARG | Concentration (μM) | 0.2 | 0.2 | 0.2 | 0.2 | 0.2 | 0.2 |
Accuracy (%) | −1.90 | −8.80 | 2.20 | −3.60 | −9.00 | 10.9 | |
Precision (%) | 2.68 | 4.11 | 2.60 | 6.16 | 2.46 | 5.73 | |
| |||||||
CIT | Concentration (μM) | 0.04 | 0.04 | 0.15 | 0.2 | 4 | 0.4 |
Accuracy (%) | −2.5 | 3.15 | −5.20 | 7.40 | 9.70 | 10.8 | |
Precision (%) | 5.46 | 6.44 | 8.66 | 2.19 | 2.38 | 4.98 | |
| |||||||
ADMA | Concentration (μM) | 0.02 | 0.02 | 0.04 | 0.05 | 0.04 | 0.04 |
Accuracy (%) | −7.50 | 4.30 | 10.10 | −6.40 | 2.05 | −12.8 | |
Precision (%) | 4.67 | 5.72 | 4.70 | 10.4 | 8.50 | 9.57 | |
| |||||||
SDMA | Concentration (μM) | 0.02 | 0.02 | 0.02 | 0.025 | 0.02 | 0.2 |
Accuracy (%) | −3.4 | −1.30 | 8.70 | 4.96 | 0.50 | 7.30 | |
Precision (%) | 5.4 | 6.33 | 2.55 | 10.8 | 8.56 | 11.2 |
B. Interday variability | |||||||
---|---|---|---|---|---|---|---|
Matrix | Mobile phase B | Cell incubation medium | Cell lysate, LGLAa | Cell lysate, HGHAb | Rat plasma | Rat urine | |
ARG | Concentration (μM) | 0.2 | 0.2 | 8 | 10 | 8 | 8 |
Accuracy (%) | 4.60 | 3.50 | −1.78 | −4.88 | 6.90 | 7.90 | |
Precision (%) | 13.6 | 11.7 | 8.35 | 8.50 | 9.27 | 5.09 | |
| |||||||
15N4-ARG | Concentration (μM) | 0.2 | 0.2 | 0.2 | 0.2 | 0.2 | 0.2 |
Accuracy (%) | 4.30 | −5.50 | 0.70 | 3.60 | −1.50 | 9.70 | |
Precision (%) | 7.97 | 6.89 | 3.59 | 5.60 | 7.11 | 4.52 | |
| |||||||
CIT | Concentration (μM) | 0.04 | 0.04 | 0.15 | 0.2 | 4 | 0.4 |
Accuracy (%) | 0.85 | 7.30 | −1.73 | 7.10 | 8.30 | 4.10 | |
Precision (%) | 6.34 | 5.51 | 9.26 | 2.32 | 3.81 | 9.60 | |
| |||||||
ADMA | Concentration (μM) | 0.02 | 0.02 | 0.04 | 0.05 | 0.04 | 0.04 |
Accuracy (%) | −3.30 | −7.00 | 1.20 | −1.12 | 3.25 | −12.1 | |
Precision (%) | 2.09 | 4.11 | 11.3 | 14.4 | 8.90 | 4.59 | |
| |||||||
SDMA | Concentration (μM) | 0.02 | 0.02 | 0.02 | 0.025 | 0.02 | 0.2 |
Accuracy (%) | −2.60 | −7.00 | −0.20 | 5.52 | 1.30 | 10.3 | |
Precision (%) | 9.74 | 4.81 | 11.1 | 9.24 | 11.4 | 10.5 |
Cell lysate obtained from EA.hy926 cells after grown in low glucose and low arginine condition
Cell lysate obtained from EA.hy926 cells after grown in high glucose and high arginine condition.