Skip to main content
. Author manuscript; available in PMC: 2012 Mar 1.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2011 Jan 14;879(7-8):467–474. doi: 10.1016/j.jchromb.2011.01.006

Table 3.

Intra- and inter- day accuracy and precision of QC samples at the lowest concentration for 5 analytes in 6 media (n=5 each). Accuracy is determined as the percentage deviation of calculated concentrations from the nominal concentrations; Precision is determined as the coefficient of variation (CV) of the 5 measurements. Additional data using higher concentrations can be seen in Supplementary Table 3.

A. Intraday variability
Matrix Mobile phase B Cell incubation medium Cell lysate, LGLAa Cell lysate, HGHAb Rat plasma Rat urine
ARG Concentration (μM) 0.2 0.2 8 10 8 8
Accuracy (%) −8.30 −1.80 8.80 6.20 −0.33 12.4
Precision (%) 3.42 11.1 3.62 4.18 2.59 1.77

15N4-ARG Concentration (μM) 0.2 0.2 0.2 0.2 0.2 0.2
Accuracy (%) −1.90 −8.80 2.20 −3.60 −9.00 10.9
Precision (%) 2.68 4.11 2.60 6.16 2.46 5.73

CIT Concentration (μM) 0.04 0.04 0.15 0.2 4 0.4
Accuracy (%) −2.5 3.15 −5.20 7.40 9.70 10.8
Precision (%) 5.46 6.44 8.66 2.19 2.38 4.98

ADMA Concentration (μM) 0.02 0.02 0.04 0.05 0.04 0.04
Accuracy (%) −7.50 4.30 10.10 −6.40 2.05 −12.8
Precision (%) 4.67 5.72 4.70 10.4 8.50 9.57

SDMA Concentration (μM) 0.02 0.02 0.02 0.025 0.02 0.2
Accuracy (%) −3.4 −1.30 8.70 4.96 0.50 7.30
Precision (%) 5.4 6.33 2.55 10.8 8.56 11.2
B. Interday variability
Matrix Mobile phase B Cell incubation medium Cell lysate, LGLAa Cell lysate, HGHAb Rat plasma Rat urine
ARG Concentration (μM) 0.2 0.2 8 10 8 8
Accuracy (%) 4.60 3.50 −1.78 −4.88 6.90 7.90
Precision (%) 13.6 11.7 8.35 8.50 9.27 5.09

15N4-ARG Concentration (μM) 0.2 0.2 0.2 0.2 0.2 0.2
Accuracy (%) 4.30 −5.50 0.70 3.60 −1.50 9.70
Precision (%) 7.97 6.89 3.59 5.60 7.11 4.52

CIT Concentration (μM) 0.04 0.04 0.15 0.2 4 0.4
Accuracy (%) 0.85 7.30 −1.73 7.10 8.30 4.10
Precision (%) 6.34 5.51 9.26 2.32 3.81 9.60

ADMA Concentration (μM) 0.02 0.02 0.04 0.05 0.04 0.04
Accuracy (%) −3.30 −7.00 1.20 −1.12 3.25 −12.1
Precision (%) 2.09 4.11 11.3 14.4 8.90 4.59

SDMA Concentration (μM) 0.02 0.02 0.02 0.025 0.02 0.2
Accuracy (%) −2.60 −7.00 −0.20 5.52 1.30 10.3
Precision (%) 9.74 4.81 11.1 9.24 11.4 10.5
a

Cell lysate obtained from EA.hy926 cells after grown in low glucose and low arginine condition

b

Cell lysate obtained from EA.hy926 cells after grown in high glucose and high arginine condition.

HHS Vulnerability Disclosure