Published in final edited form as: J Magn Reson Imaging. 2010 Jul;32(1):173–183. doi: 10.1002/jmri.22213

Table 1.

Acquisition Parameters for MR Sequences

Imaging
Parameters
FSE-
Cube
MENSA IDEAL-
GRASS
IDEAL-
SPGR
COSMIC VIPR-
bSSFP
TR/TE (ms) 2217/23.6 16.7/5.7 12.4/5.3 12.4/5.3 4.6/1.3 3.6/0.3
Flip Angle 90° 30° 50° 14° 35° 15°
Bandwidth (kHz) 31.2 31.2 31.2 31.2 83.3 125
Field of View (cm) 15 15 15 15 15 15
Matrix 256 × 256 256 × 256 384 × 224 384 × 224 256 × 256 384 × 384
Slice Thickness (mm) 0.7 0.7 1.0 1.0 0.7 1.2*
Voxel Volume (mm3) 0.24 0.24 0.26 0.26 0.24 0.18*
Scan Time (min) 5 5 5 5 5 5
*

Three-slice averages in multiple planes were used to obtain 0.39×0.39 in-plane resolution and 1.2mm sections.

FSE-Cube = fast spin-echo Cube,

MENSA = multi-echo in the steady-state acquisition,

IDEAL-GRASS = iterative decomposition of water and fat with echo asymmetry and least-squares estimation combined with unspoiled gradient recalled-echo acquired in the steady-state,

IDEAL-SPGR = iterative decomposition of water and fat with echo asymmetry and least-squares estimation combined with spoiled gradient-echo,

COSMIC = coherent oscillatory state acquisition for the manipulation of image contrast,

VIPR-bSSFP = vastly undersampled isotropic projection reconstruction balanced steady-state free precession,

TR: repetition time,

TE: echo time.