Skip to main content
. 2011 Mar 11;38(4):1877–1887. doi: 10.1118/1.3553405

Figure 3.

Figure 3

(a) The relative acoustic pressure squared distribution measured by a needle hydrophone (0.2 mm in diameter) in a tank filled with degassed water (applied electrical power: 1 W, frequency: 2.25 MHz). (b) The corresponding simulated relative acoustic pressure squared distribution.