Published in final edited form as: Magn Reson Med. 2008 Nov;60(5):1122–1134. doi: 10.1002/mrm.21737

Table 1.

The values of the pre-calibrated and self-calibrated relative amplitudes for the data shown in Figure 49. Chemical shift frequencies of the peaks relative to water are labeled for 3T.

Data α1 α2 α3 α4 α5 α6
420Hz 318Hz −94Hz 472Hz 234Hz 46Hz
pre-calibration, 16-echo IDEAL, peanut oil 0.62 0.15 0.10 0.06 0.03 0.04
Fig. 4, 3-pt 1.5T FSE peanut oil, self calibration 0.71 0.21e(−j0.4π) 0.09 0 0 0
Fig. 5, 3-pt, 3T FSE Knee, self calibration 0.78 0.15e(−j0.4π) 0.08 0 0 0
Fig. 6, 3-pt, 3T FSE Ankle, self calibration 0.74 0.16e(−j0.5π) 0.10e(j0.1π) 0 0 0
Fig. 6, 3-pt 1.5T FSE Spine, self calibration 0.74 0.18e(−j0.4π) 0.08 0 0 0
Fig. 7, 3-pt, 1.5T SPGR Ankle, center echo: 5π/2, self calibration 0.68 0.23e(−j0.1π) 0.08e(j0.3π) 0 0 0
Fig. 7, 3-pt, 1.5T SPGR Breast, center echo: 3π/2, self calibration 0.85 0.08e(j0.9π) 0.08e(j0.2π) 0 0 0
Fig. 8, 6-pt, 1.5T liver, self calibration 0.75 0.17 0.08 0 0 0
Fig. 9, 6-pt, 1.5T liver, patient 1, self calibration 0.72 0.20 0.08 0 0 0
Fig. 9, 6-pt, 1.5T liver, patient 2, self calibration 0.76 0.16 0.08 0 0 0