Skip to main content
. Author manuscript; available in PMC: 2012 Apr 1.
Published in final edited form as: J Magn Reson Imaging. 2011 Apr;33(4):792–802. doi: 10.1002/jmri.22507

Table 1.

3D MRSI acquisition protocols that were considered in this study

STANDARD OCTAGONAL AUTOSAT AUTOSAT+
OCTAGONAL
Fixed ROI-edge sat bands
Manually prescribed sat bands
Octagonal selection
Automatically prescribed sat bands
FOV (cm) 16×16×16 16×16×16 18×18×16 18×18×16
Lactate Editing
TR 1104 1134 1500 1500
Tacq (min) 9 9 6.5 6.5
Slices in PRESS box 4 4 6 6

STANDARD = standard lactate edited MRSI, OCTAGONAL = lactate-edited MRSI with octagonal selection, AUTOSAT = MRSI with automatically generated sat band placement, AUTOSAT+OCTAGONAL = combination of automatic sat band placement and octagonal selection