Table 3.
Evaluation of PCR on RDT on 12 RDT brands
| Results for P. falciparum with 21/μl PCR on whole blood Ct 33.52 | Results for P. falciparum with 830/μl PCR on whole blood Ct 26.86 | |||||
|---|---|---|---|---|---|---|
| RDT name | blood volume | Ag detection | DNA detection (Ct) | Ag detection | DNA detection (Ct) | RDT fragment used |
| OptiMAL | 10 μl | n | 34,67 | p | 30,12 | 1/3 NC |
| SDFK50 | 5 μl | p | 35,32 | p | 30,36 | 1/2 NC + C |
| SDFK60 | 5 μl | p | 35,47 | p | 30,20 | 1/2 NC + C |
| Pan MAL card | 5 μl | n | 35,77 | p | 30,49 | 1/2 NC |
| First response* | 5 μl | p | 35,88 | p | 31,48 | 2/3 S/C |
| ICT | 5 μl | p | 36,74 | p | 30,18 | 1/2 NC |
| ParaHIT | 8 μl | p | 37,56 | p | 33,20 | 1/2 C |
| Binax | 15 μl | p | 37,77 | p | 30,19 | 1/2 NC |
| Core* | 5 μl | p | 38,82 | p | 32,83 | 2/3 S/C |
| Hexagon | 5 μl | p | 39,58 | p | 34,08 | 1/2 NC |
| Paracheck* | 5 μl | p | 40,53** | p | 37,33 | C |
| Carestart* | 5 μl | p | 0 | p | 39,21 | 2/3 S/C |
-Antigen detection is considered positive (p) as test line(s) was (were) visible and negative (n) if no test line was visible.
-DNA detection by real-time PCR on RDT is expressed in Cycle threshold (Ct)-values and RDTs were ranked in order of increasing Ct-values.
-RDT fragments used were either one third (1/3) or half (1/2) of the proximal part of the nitrocellulose (NC) strip or (two third of) the sample (S) and/or conjugate (C) pad (see Figure 2).
* RDTs with a plastic seal covering ik the nitrocellulose strip
** In the first test, no signal was seen (Ct-value 0), only by repetition of the DNA extraction a positive signal (Ct-value 40.53) was observed.