Table 2.
Variable | Level + | Level − | Motivation |
---|---|---|---|
Position of
vacuum peak |
1 mm from ESI inlet |
1 mm from MS inlet |
Previous experiments showed that these 2 parameters could have a positive impact |
ESI Gap | 9 mm | 6 mm | As no improvement was found at 3mm with the amplifier during previous experiment, larger space was explored |
ESI Voltage | 3200 V | 2800 V | Range that showed good performance in previous experiments |
N2 Temperature | 100°C | 25°C | Hot N2 can improve desolvation |
Air amplifier
pressure |
45 PSI | 30 PSI | Amplifier pressure range |
Composition | 100% H20 | 25% H20/75% ACN |
Test the hypothesis that the air amplifier can improve desolvation of aqueous solvents. |
Air amplifier
radial alignment |
Centered | 0.75 mm offset |
Previous experiments performed earlier showed that some |
ESI Tip size | 100 μm | 75 μm | As previous experiments showed no overall signal improvement with air amplifier. This is to test the hypothesis that better results can be obtained with larger tips |
Flow | 2 μL/min | 1 μL/min |