. 2010 Aug 19;53(5):359–365. doi: 10.1007/s00234-010-0753-y

Table 3.

Detailed results for relative in-stent signal (RIS) and artificial lumen narrowing (ALN)

  Average RIS (%) for CE-MRA Average RIS (%) for 3D-TOF p value (RIS CE-MRA vs. 3D-TOF) ALN (%) for CE-MRA ALN (%) for 3D-TOF
Wal8 60.0 13.2 0.0006 25.0 60.0
Wal6 54.1 9.1 0.0005 30.0 100.0
Wal4 16.3 4.7 0.0003 75.0 100.0
Wallstent® average 43.5 9.0 43.3 86.7
Zil8 100.8 98.7 0.3154 8.3 30.0
Zil6 85.4 59.7 0.0158 20.0 25.0
Zil4 69.4 71.1 0.3763 37.5 20.0
Zilver® average 85.2 76.5 21.9 25.0
Acc8 80.8 91.5 0.0508 0.0 20.0
Acc6 89.6 73.4 0.0439 0.0 25.0
Acc4 67.0 72.7 0.2691 25.0 20.0
Acculink® average 79.2 79.2 8.3 21.7
Pre8 56.3 49.2 0.0647 0.0 30.0
Pre6 37.5 42.8 0.0161 30.0 37.5
Pre4 26.9 39.7 0.0633 37.5 20.0
Precise® average 40.3 43.9 22.5 29.2
Xac8 10.7 3.9 0.0126 91.7 100.0
Xac6 8.4 2.8 0.0059 100.0 100.0
Xact® average 9.6 3.3 95.8 100.0