Skip to main content
. 2011 Mar 23;6:23. doi: 10.1186/1746-1596-6-23

Table 2.

Results by device type

Device Efficacy Manual Powered p-value
Able to acquire biopsy core specimen 98.0% 90.4% 0.205

Mean number attempts for core acquisition 1.5 ± 0.8 1.2 ± 0.5 0.112

Mean time core acquisition (seconds) 203.1 ± 149.5 102.1 ± 85.4 0.000*

Patient Pain Scores Manual Powered p-value

Mean VAS needle insertion (0-10) 3.2 ± 2.9 3.1 ± 3.1 0.900

Mean VAS overall (0-10) 3.5 ± 2.3 3.8 ± 2.8 0.623

Pain-free patients 24 hours post procedure 33.3% 66.7% 0.003*

Pain-free patients 7 days post procedure 76.1% 82.6% 0.607

Satisfaction & Complications Manual Powered p-value

Operator satisfaction (0-10) 7.8 ± 2.1 8.1 ± 1.7 0.488

Patient satisfaction (0-10) 9.3 ± 1.6 9.4 ± 1.2 0.778

Device related complications 0% 3.8% 0.495

Core Biopsy Pathology Assessment Manual Powered p-value

Mean length (mm) 11.0 ± 5.5 13.3 ± 6.6 0.368

Mean width (mm) 1.3 ± 0.5 1.7 ± 0.5 0.123

Mean volume (mm3) 20.4 ± 9.0 36.8 ± 21.2 0.039*

General quality excellent/good 80.0% 66.7% 0.628

Medullary quality excellent/good 70.0% 55.6% 0.650

Overall quality rated adequate 70.0% 77.8% 1.000

*denotes statistical significance