Published in final edited form as: J Magn Reson Imaging. 2011 May;33(5):1128–1135. doi: 10.1002/jmri.22537

Table 1.

Measured PRF temperature sensitivity coefficients from three calibrations at 1.5T

Chemical Shift TSC [ppm/°C] (R2)
Water −1.07 × 10−2 ± 1.0 × 10−4 (0.981) −1.11 × 10−2 ± 2.4 × 10−4 (0.974) −0.98 × 10−2 ± 1.3 × 10−4 (0.969)
Bulk Methylene −0.20 × 10−2 ± 1.0 × 10−4 (0.438) −0.24 × 10−2 ± 1.5 × 10−4 (0.738) −0.12 × 10−2 ± 1.1 × 10−4 (0.440)
Difference −0.87 × 10−2 ± 2.7 × 10−4 (0.942) −0.87 × 10−2 ± 4.0 × 10−4 (0.961) −0.85 × 10−2 ± 3.2 × 10−4 (0.920)