Skip to main content
. Author manuscript; available in PMC: 2011 Oct 1.
Published in final edited form as: Magn Reson Med. 2010 Oct;64(4):1171–1181. doi: 10.1002/mrm.22434

Table 1.

Imaging Parameters for Experimental Studies*

Imaging parameters Phantom experiments parameters
In vivo parameters, 8 × calves
Full ref. Circular ref. CAPR ref.
TR/TE (msec) 5.9/2.7 5.9/2.7 5.9/2.7 5.9/2.7
Flip angle (degrees) 30 30 30 30
Bandwidth (± kHz) 62.5 62.5 62.5 62.5
Sampling matrix (Nx × Ny × Nz) 400 × 320 × 132 400 × 320 × 132 400 × 320 × 132 400 × 320 × 132
FOV (X × Y × Z cm3) 40 × 32 × 13.2 40 × 32 × 13.2 40 × 32 × 13.2 40 × 32 × 13.2
Spatial resolution (X × Y × Z cm3) 1 × 1 × 1 1 × 1 × 1 1 × 1 × 1 1 × 1 × 1
CAPR specific parameters
 Center size 42,240 32,788 400 400
 View sharing N1 N1 N4 N4
 CAPR HD undersampling factor (RHD) 0.78 1 1.98 1.95 1.91 1.82 1.82
 2D SENSE acceleration (R) 1 2 4 8 8
 Rnet CAPR = RHD × R 0.78 1 1.98 3.9 7.64 14.56 14.56
 Max CAPR undersampling (RUS) 3.73 3.5 3.12 2.59 2.59
 Rnet Max CAPR = RHD × R × RUS 7.39 13.65 23.84 37.7 37.7
Temporal footprint (sec)
 CAPR 249.2 193.4 101.8 53.6 29.5 17.6 17.6
 Max CAPR 26 14 8 4.9 4.9
Frame time (sec)
 CAPR 249.2 193.4 26 14 8 4.9 4.9
 Max CAPR 26 14 8 4.9 4.9
 Associated figure 3 3 4, 5, 7, 8
*

The Full ref. column is for the case in which a full rectangular region in kY-kZ space of Ny × Nz points is sampled. For the Circular ref. case, the corners of kY-kZ space are not sampled, as discussed in the text. The parameter “Nm” for view-sharing indicates that “m” vane sets are used. The set of four values listed in the CAPR ref. column for each of RHD, R, RUS, Rnet CAPR, Rnet Max CAPR, temporal footprint, and frame time is for respective SENSE accelerations of R = 1, 2, 4, 8.