Skip to main content
. 2011 Feb 22;286(19):16546–16554. doi: 10.1074/jbc.M110.215319

TABLE 3.

Summary of HTS quality control parameters for 96-well plates

Parameter Accumulation
Degradation
UbG76V-GFP reporter ODD-Luc reporter UbG76V-GFP reporter ODD-Luc reporter
S/Ba 3.4 11 2.6 5.8
CVb 3.3% (P), 1.9% (N) 7.3% (P), 4.5% (N) 7.4% (P), 7.3% (N) 8.2% (P), 7.6% (N)
Z′-factorc 0.67 0.7 0.5 0.57

a S/B, signal/base line ratio.

b CV, coefficient of variation ((S.D./mean) × 100).

c Z′-factor = 1 − ((3 × (σp + σN))/(μP − μN)), where σ is S.D. and μ is the mean for positive (P) and negative (N) controls. The positive control was MG132 (2 μm), and the negative control was DMSO (0.8%).