TABLE 1.
Reproducibility of the BTUB FRET PCR assay
| Technician | Standard (known concn) | Mean calculated concn (SD)a,b | Mean Ct (SD)a,c |
|---|---|---|---|
| 1 | 1 (84) | 88.1 (21.4) | 30.2 (0.40) |
| 2 (42) | 38.8 (5.65) | 31.5 (0.23) | |
| 3 (21) | 19.4 (3.81) | 32.6 (0.35) | |
| 4 (10.5) | 12.2 (1.50) | 33.3 (0.21) | |
| 5 (5.25) | 6.50 (3.30) | 34.5 (0.78) | |
| 6 (1.06) | 0.84 (0.03) | 37.6 (0.05) | |
| 2 | 1 (84) | 91.1 (21.6) | 31.0 (0.45) |
| 2 (42) | 45.6 (10.5) | 32.4 (0.44) | |
| 3 (21) | 22.3 (3.80) | 33.5 (0.33) | |
| 4 (10.5) | 10.6 (2.47) | 34.7 (0.40) | |
| 5 (5.25) | 4.48 (0.95) | 36.4 (0.39) | |
| 6 (1.06) | 1.62 (1.10) | 38.4 (1.30) | |
| Pooled means | 1 (84) | 89.6 (21.5) | 30.6 (0.43) |
| 2 (42) | 40.7 (8.10) | 31.9 (0.33) | |
| 3 (21) | 20.8 (3.80) | 33.1 (0.34) | |
| 4 (10.5) | 11.4 (1.90) | 34.1 (0.30) | |
| 5 (5.25) | 5.40 (2.10) | 35.5 (0.60) | |
| 6 (1.06) | 1.23 (0.60) | 38.0 (0.70) |
All values are means based on five replicates analyzed for each technician and 10 replicates analyzed for the pooled means.
T. vaginalis organisms per PCR, generated by the LightCycler utilizing Ct and the known concentration of the standards.
Ct was generated based on the cycle number where the LightCycler software interpreted a logarithmic increase in fluorescence of the given sample.