Skip to main content
. 2011 Jun;52(6):1200–1210. doi: 10.1194/jlr.M013144

TABLE 1.

Accuracy and precision of D-4F plasma measurements

Peptide Added (ng/ml) 1,000 200 40 10 1
Values obtained from quadruplicates (ng/ml) 925 206 41.8 9.31 1.32
965 192 34.5 8.46 1.17
961 187 35.4 7.24 0.931
1090 185 35.0 6.75 1.03
Mean ± SD 985.3 ± 72 192.5 ± 9.47 36.68 ± 3.43 7.94 ± 1.16 1.11 ± 0.169
95% CI (870.5–1100) (177.4–207.6) (31.21–42.14) (6.09–9.79) (0.84–1.38)
CV% 7.3% 4.9% 9.4% 14.6% 15.2%

A plasma sample from an apoE null mouse that had not received peptide was spiked with D-4F to give a plasma concentration of 1,000, 200, 40, 10, or 1 ng/ml. The plasma was extracted and the D-4F concentration was determined in quadruplicate as described in Materials and Methods.