Skip to main content
. Author manuscript; available in PMC: 2011 May 10.
Published in final edited form as: J Pharm Biomed Anal. 2010 Oct 13;54(3):433–438. doi: 10.1016/j.jpba.2010.08.025

Table 1.

System suitability test results.

Parameters Specifications Day 1 Day 2 Day 3
Retention time (% RSD) ≤2.0 0.65 0.75 0.12
Capacity factor (k′) >3.0 5.64 5.70 5.63
Area (% RSD) ≤2.0 0.95 0.84 0.42
Plates (column) >3000     43,725     30,900     28,200    
USP tailing <1.5 1.24 1.43 1.39