Skip to main content
. 2010 Jul-Dec;5(2):107–117.

Table 3.

Inter-day assay precision and accuracy of the HPLC method for the determination of NAG*

Nominal concentration (µg/ml) Calculated concentration (µg/ml) Precision RSD (%) Accuracy error
1 0.935 6.46 2.41
2 2.009 3.53 -0.454
4 4.07 2.94 -1.925
6 5.93 1.59 1.215
8 8.17 1.01 -2.146
10 9.88 0.42 1.188
*

Validation study, n=5 days, five replicates per day