Quality control program for a computed radiography system

Frequency Tasks Responsibility
Daily System inspection for physical defects Radiographer
Physical inspection of display devices
Secondary erasure of imaging plates
Verification of system interface/network
Weekly Verification of displayed images Radiographer
Phantom image quality control testing
- Image quality
- Artefacts
Monthly Inspect and clean image receptors Radiographer
Review image rejection rate
QC review for ‘out-of-tolerance’ issues
Semi-annually/Annually Evaluate image quality and patient dose Medical physicist
Acceptance tests to re-establish baseline value
Review for:
- Patient exposure trends
- Retake activity
- QC records
- Service history