MQSA inspection results for selected technical parameters: January to December, 1995.

Unless otherwise indicated, tabulated values are: N / Mean / SD Hospitals Private Practice Breast Clinics Other facilities ALL
Mean Glandular Dosea (mGy) 5715/1.50/0.42 4959/1.51/0.42 730/1.47/0.39 293/1.58/0.44 11697/1.51/0.42
kVp 5732/25.9/1.2 4977/26.0/1.4 734/25.9/1.0 293/25.9/1.1 11736/26.0/1.2
HVL (mm Al) 5732/0.32/0.03 4977/0.33/0.05 734/0.33/0.02 293/0.32/0.02 11736/0.33/0.04
Phantom Image Background OD 5732/1.43/0.22 4977/1.42/0.23 734/1.44/0.21 293/1.41/0.23 11736/1.42/0.23
Fibersb 5729/4.6/0.6 4923/4.5/0.6 731/4.6/0.6 293/4.5/0.7 11676/4.5/0.6
Specksb 5729/3.7/0.5 4923/3.7/0.5 731/3.7/0.5 293/3.7/0.6 11676/3.7/0.5
Massesb 5729/3.6/0.6 4923/3.6/0.6 731/3.7/0.5 293/3.6/0.6 11676/3.6/0.6
Total Net Scorec 5705/11.5/1.1 4955/11.4/1.1 732/11.6/1.0 290/11.4/1.1 11682/11.5/1.1
Film Processing: % < 80d 4.5 (80/1777) 3.4 (60/1742) 1.8 (2/109) 8.0 (7/88) 4.0 (149/3717)
Darkroom Fog: % > 0.05 ODe 10.5 (352/3365) 12.1 (381/3156) 8.8 (29/328) 8.8 (16/181) 11.1 (778/7030)
% Phantom Images with artifacts 62.2 (3565/5729) 57.1 (2812/4923) 56.9 (416/731) 63.5 (186/293) 59.8 (6979/11676)
a

Dose was calculated from data acquired by the MQSA inspector.

b

Object score is reported without artifact subtraction.

c

Score includes artifact subtraction.

d

Percentage of standard-cycle film processors with a speed less than 80, as determined by the STEP method.

e

Percentage of facilities found to have darkroom fog greater than 0.05 net optical density.