. 2007 Apr 1;3(2):e35. doi: 10.2349/biij.3.2.e35

Table 2.

MQSA inspection results for selected technical parameters: January to September 2006

Unless otherwise indicated, tabulated values are:N/Mean/ SD Hospitals Private Practice Breast Clinics Other facilities ALL Digital unitsa Screen-FilmUnits
Mean Glandular Doseb (mGy) 4776/1.78/0.33 3681/1.79/0.35 806/1.81/0.33 359/1.82/0.31 9776/1.78/0.34 890/1.63/0.39 8542/1.80/0.33
kVp 4271/25.5/0.7 3328/25.5/0.7 653/25.5/0.7 334/25.5/0.7 8586/25.5/0.7 N/A 8586/25.5/0.7
HVL (mm Al)c 2227/0.33/0.02 1814/0.33/0.02 418/0.32/0.02 150/0.33/0.02 4609/0.33/0.02 N/A 4609/0.33/0.02
Phantom Image Background OD 4276/1.85/0.22 3328/1.85/1.20 653/1.89/0.20 334/1.85/0.22 8591/1.85/0.21 N/A 8591/1.85/0.21
Fibersd 4614/5.1/0.6 3573/5.0/0.6 771/5.1/0.6 351/5.0/0.6 9135/5.1/0.6 892/5.3/0.6 8568/5.0/0.6
Specksd 4614/3.9/0.3 3573/3.9/0.4 771/3.9/0.3 351/3.9/0.4 9135/3.9/0.3 892/4.0/0.3 8568/3.9/0.3
Massesd 4614/3.9/0.5 3573/3.9/0.5 771/4.0/0.4 351/3.9/0.4 9135/3.9/0.4 892/4.3/0.4 8568/3.9/0.4
Total Net Scoree 4614/12.4/1.1 3725/12.4/1.1 770/12.6/1.1 351/12.3/1.1 9309/12.4/1.1 891/13.5/0.9 8567/12.3/1.1
Film Processing: % < 80f 0.1 (2/2852) 0.0 (0/2389) 0.0 (0/358) 0.0 (0/262) 0.0 (2/5861) 0.0 (0/166)h 0.0 (2/5761)
Darkroom Fog: % > 0.05 ODg 4.7 (127/2725) 4.0 (92/2325) 1.3 (4/304) 4.3 (10/233) 4.2 (233/5587) 3.2 (7/220)h 4.2 (235/5565)
% Phantom Images with artifacts 69.9 (3226/4614) 69.9 (2498/3573) 69.1 (533/771) 69.8 (245/351) 69.9 (6502/9307) 29.9 (267/892) 73.2 (6270/8568)
a

18 facilities that were inspected between August (during which MQSA inspections of digital equipment began) and December 2005 were also included in the population of digital sites.

b

Dose is that reported by the Medical Physicist.

c

In mid-2006 FDA ceased equipment radiation measurements for HVL, hence the sample size is smaller than those for the remaining tabulated parameters.

d

Object score is reported without artifact subtraction.

e

Score includes artifact subtraction.

f

Percentage of standard-cycle film processors with a speed less than 80, as determined by the STEP method.

g

Percentage of facilities found to have darkroom fog greater than 0.05 net optical density.

h

If the facility also operated at least one screen-film mammography unit then the film processing quality and darkroom fog were evaluated.