Figure 5.

Figure 5

Predicted effects of two major parameters on predicted DFT based on general linear model. Left panel: effect of increase in Metric A (worsening of alignment of shock vector with ventricular myocardial center of mass) to increase predicted DFT. Right panel: effect of increase in electrode length to decrease predicted DFT by smaller increment.