Skip to main content
. Author manuscript; available in PMC: 2012 Jan 1.
Published in final edited form as: J Liq Chromatogr Relat Technol. 2011;37(7):491–499. doi: 10.1080/10826076.2011.556965

Fig.3.

Fig.3

Analytical HSCCC separation of the crude sample obtained from C. militaris after cleaning-up by 732 cation-exchange resin. Solvent system: n-hexane–n-butanol–methanol–water (23:80:30:155, v/v/v/v); stationary phase: upper phase; mobile phase: lower phase; flow rate: 1.0 ml/min; evolution speed: 1600 rpm; sample: 4.2 mg dissolved in 2 ml of the lower phase; retention of the stationary phase: 49%; detection wavelength: 254 nm. Peak c: cordycepin.