Review of biliary stent trimming

Auther Total No. of cases SEMS No. of case Setting of APC system Procedural time Success rate
Demarquay et al[4] 3 Uncovered Wallstent 3 85 W, coag. 0.8 L/min NA 100% (3/3)
Vanbiervliet et al[10] 24 Uncovered Wallstent 23 70-80 W, coag. 0.8 L/min < 30 min 95.7% (22/23)
Covered Wallstent 1 70-80 W, coag. 0.8 L/min < 30 min 0% (0/1)
Guda et al[11] 2 Uncovered Wallstent 1 80 W, coag. 2.0 L/min NA 100% (1/1)
Uncovered Wallstent 1 60 W, coag. 2.0 L/min NA 100% (1/1)
Rerknimitr et al[12] 2 Covered Wallstent 1 60 W, coag. 0.8 L/min 20 min 100% (1/1)
Uncovered Wallstent 1 70 W, coag. 0.8 L/min NA 100% (1/1)
Christiaens et al[13] 2 Uncovered SEMS (nitinol) 1 60 W, coag. 1.8 L/min NA 100% (1/1)
Uncovered SEMS (nitinol) 1 60 W, coag. 1.8 L/min NA 100% (1/1)
Present study 8 Covered Wallstent 3 60-80 W, coag. 1.5 L/min < 20 min 100% (3/3)
Uncovered Wallstent 1 60-80 W, coag. 1.5 L/min 12 min 100% (1/1)
Niti-S Biliary ComVi Stent 4 60-80 W, coag. 1.5 L/min < 20 min 100% (4/4)

NA: Not available; SEMS: Self-expandable metallic stent; APC: Argon plasma coagulation; coag.: Coagulation mode.