Abstract
BACKGROUND
Use of the QuantiFERON®-TB Gold In-Tube assay (QFT-GIT) in remote areas is limited by the need to incubate blood samples within 12 h of collection. PortaTherm™ is a portable, electricity-free, phase-change incubator previously used for field collection of microbiological samples.
OBJECTIVE
To determine whether the PortaTherm can be used for the reliable incubation of QFT-GIT samples, thus enabling QFT-GIT use in settings distant from laboratory facilities.
METHODS
In a prospective comparative study in Peru, blood samples were collected from 50 participants and processed in three parallel QFT-GIT tests per participant; two were incubated in a conventional incubator; the third was incubated in the PortaTherm.
RESULTS
All 150 QFT-GIT tests gave definitive results, and for 46 of the 50 participants all three tests were concordant, eight of which were positive. Four participants had one discordant result: two due to discordance of a conventional incubator QFT-GIT result, and two due to discordant PortaTherm QFT-GIT results.
CONCLUSION
The QFT-GIT inter-incubator variability between the PortaTherm and conventional incubator was no greater than the intra-incubator variability for the conventional incubator, indicating that the PortaTherm is a suitable tool for incubating QFT-GIT whole blood samples in remote settings where access to a laboratory or electricity is limited.
Keywords: incubator, PortaTherm, QuantiFERON®-TB Gold In-Tube, tuberculosis, inter-assay variability
IN INDUSTRIALIZED COUNTRIES, interferon-gamma release assays (IGRAs) are increasingly being used in preference to the tuberculin test for the identification of latent tuberculosis infection (LTBI) due to greater specificity (IGRAs are unaffected by prior subject bacille Calmette-Guérin vaccination).1,2 In the US Food and Drug Administration-approved QuantiFERON®-TB Gold In-Tube assay (QFT-GIT, Cellestis Limited, Carnegie, VIC, Australia), whole blood samples taken into three separate tubes are exposed to either no stimulation (nil tube, negative control), maximal stimulation (mitogen tube, positive control) or stimulation with tuberculosis (TB) specific antigens (TB antigen tube). In each tube, significant differential stimulation of T-lymphocytes in the whole blood occurs only during controlled incubation at 37°C, after which the levels of interferon-gamma (IFN-γ) released into harvested plasma are detected by enzyme-linked immunosorbent assay (ELISA). Measured absorbance values are converted to international units of concentration using software provided by the manufacturer (QuantiFERON-TB In-Tube V2.50), and results are reported as positive, negative or indeterminate according to predetermined cut-off criteria.
In addition to its use for the clinical care of individuals, QFT-GIT testing can be used for surveying populations and monitoring TB infection prevalence over time.3-5 However, because the blood samples used in performing the QFT-GIT assay require incubation within 12 h of collection, the test cannot be employed in remote areas or at sites distant from laboratory facilities. In the present study, we report the evaluation of a low-cost, portable incubator with the potential to circumvent this practical limitation of the QFT-GIT assay, allowing for remote testing in the field.
THE PORTATHERM PHASE-CHANGE INCUBATOR
Certain substances change phase from liquid to solid (and vice versa) at a constant phase-change temperature until the entire body of the substance has changed phase (e.g., as melting ice maintains water at 0°C until it has melted completely).6,7 The PortaTherm™, developed in the D-Lab at Massachusetts Institute of Technology (Cambridge, MA, USA), is a portable incubator that exploits this phenomenon of phase-change temperature stability utilizing material with a phase-change temperature of 37°C. The white solid material is sealed inside transparent polythene or plastic packages and heated in a water bath (for example, a pan on an open fire) until melted (the liquid form is transparent), and then allowed to cool until the first indication of cloudiness due to white precipitates of solid material is visible. At this point, the packages are placed in the insulated container (a styrofoam/polystyrene box or commercially available picnic cooler), and if the PortaTherm incubator remains unopened, a temperature of 37°C can be maintained for at least 4 days with adequate insulation (Jesus Rojas/Mark Brady, unpublished data).
The PortaTherm was initially designed for the incubation of water samples for microbiological testing, and has previously been used successfully for field collection of pediatric blood cultures for Streptococcus pneumoniae and gastric aspirates for Helicobacter pylori (unpublished data). If the PortaTherm can be used to incubate QFT-GIT samples, then QFT testing of populations living in settings remote from laboratories with a conventional incubator and/or without an electricity supply would be greatly facilitated. To evaluate this potential utility, we compared the results of the QFT-GIT assay using blood samples incubated in a conventional laboratory incubator and the PortaTherm.
METHODS
Study design
A comparative study of the results of three parallel QFT-GIT assays was performed utilizing the PortaTherm phase-change incubator and a conventional incubator (NU-4850, NuAire, Plymouth, MN, USA). Two QFT-GIT assays were performed in the same conventional incubator to control for any inherent inter-assay variability not attributable to the use of the PortaTherm.
Setting and study population
From December 2008 to March 2009, 50 volunteer participants were recruited from staff at the research laboratories of Universidad Peruana Cayetano Heredia (Lima, Peru) and the Infectious Diseases Unit of Hospital Nacional Dos de Mayo, a government hospital in Lima, Peru.
Field methods
Inclusion criteria required participants to be at least 18 years of age and not undergoing treatment for TB. After giving written informed consent, participants answered a short questionnaire about TB exposure and a 9 ml sample of blood was drawn to perform three QFT-GIT tests. Each test requires three separate 1 ml vacuum collection tubes: a nil control containing no antigen, a positive control containing the mitogen phytohemagglutinin and the TB-specific tube containing the Mycobacterium tuberculosis antigens, early secreted antigenic target-6 (ESAT-6), culture filtrate proteins 10 (CFP-10) and TB7.7. Prior to incubation the blood samples were kept at room temperature for no more than 8 h.
Laboratory methods
After transfer to the laboratory the first two sets of QFT-GIT tubes for each patient were incubated in the conventional incubator (NU-4850 #1 and #2), and the third set was incubated in the PortaTherm. The PortaTherm was prepared as described above. Briefly, packages containing the phase-change material were heated in a hot water bath at 40°C until the material was observed to completely change from solid to liquid state. The packages were then dried and carefully placed along the sides and bottom of a standard (0.38 m × 0.38 m × 0.39 m; 6.5 cm thick) Styrofoam polystyrene insulation box (Figure 1). A test-tube rack containing the three tubes for one QFT-GIT test from each subject was placed in the center of the incubator, and surrounded and covered with further packages of phase-change material, placed so as to minimize empty air space within the box, which was filled to the top and covered with a well-fitted Styrofoam lid.
Figure 1.

PortaTherm™ phase-change incubator. A. Empty polystyrene box. B. Samples covered with packets of phase change material.
All tubes were incubated for 20 h in accordance with the manufacturer’s instructions; during this time, the internal temperature of the conventional incubator and the PortaTherm were recorded using digital temperature monitors (Smart Button™ Temperature Logger, ACR Systems, Surrey, BC, Canada) with range −10°C to +85°C and precision ±0.5°C.
Following incubation, the QFT-GIT IFN-γ assay was performed according to the manufacturer’s instructions (Cellestis Ltd); after centrifugation and harvesting of plasma, each patient’s three assays were run on the same ELISA plate.
Data analysis
The null hypothesis was that any difference between PortaTherm QFT-GIT and either of the NU-4850 QFT-GIT results would be no greater than the difference between the two NU-4850 QFT-GIT results. As non-indeterminate QFT-GIT results are reported to patients as positive or negative, the principal analysis compared this dichotomous output of QFT-GIT between the PortaTherm and NU-4850 incubators. Quantitative measurements of IFN-γ levels for the three tubes (nil, mitogen and TB antigens) are determined from absolute assay optical density (OD) values. As these quantitative data are available for all three tubes, the data were also compared. Data were analyzed using STATA 10.0 (Stata Corporation, College Station, TX, USA).
The paired t-test was used to test the differences between the mean OD values of the three tubes (nil, mitogen and TB antigens) for the three different incubator comparisons (NU-4850 #1/NU-4850 #2, NU-4850 #1/PortaTherm and NU-4850 #2/PortaTherm). The null hypothesis was that the mean of the difference between the incubators is zero (5% level of significance). Normality of the differences between the three measured values (nil, mitogen and TB antigens) was verified by the Shapiro-Wilk test.
Ethical review
The research protocol and consent form were reviewed and approved by the Institutional Ethics Committee at the Universidad Peruana Cayetano Heredia, Lima, Peru. Written informed consent was obtained from all 50 participants. Participants were informed of the results of their QFT-GIT tests performed in the conventional incubator.
RESULTS
Participant characteristics
The 50 participants were 20 research laboratory scientists from Universidad Peruana Cayetano Heredia and 30 health care personnel from the Infectious Diseases Service of the Hospital Nacional Dos de Mayo. The mean age was 29.2 years (range 20–52), and 31 were female. Only two of the 20 laboratory scientists reported known contact with a TB patient in the preceding 3 months, neither for more than 2 months. In contrast, all 30 health care personnel reported recent contact with TB patients (between 0 and 168 h prior to enrolment; mean ± standard deviation [SD] 11.2 ± 34.35 h), which was generally of prolonged duration (27.0 ± 32.46 months, range 1 day–120 months). One participant was known to be human immuno-deficiency virus infected.
Incubator temperature logs
In each run, the PortaTherm and the NU-4850 NuAire incubator maintained the target temperatures of 37°C ± 1°C throughout the 20-h incubation period (Figure 2).
Figure 2.
Smart-button™ temperature log data for the NU-4850 NuAire and PortaTherm incubators. Continuous lines show PortaTherm data, dotted lines show NU-4850 data. Each figure represents a single day’s Quantiferon run with duplicate measurements for each incubator.
Results of the QFT-GIT assay
None of the 150 QFT-GIT assays gave an indeterminate result.
Dichotomous analysis
For 46 of the 50 participants, all three QFT-GIT tests gave concordant positive/negative results (Figure 3). Thus, 8/46 participants were QFT-GIT-positive and 38 were QFT-GIT-negative; QFT-GIT positivity was much more frequent in health care workers (7/27, 25.9%) than research laboratory staff (1/19, 5.3%). Disagreement between the two assays run in the NU-4850 incubator occurred twice and between the PortaTherm and each of the NU-4850 incubator trials once each (Figure 3, raw data in Table).
Figure 3.

Venn diagram indicating between-incubator concordance for QFT-GIT positive/negative results (N = 50 participants). QFT-GIT = QuantiFERON®-TB Gold In-Tube.
Table.
Absolute IFN-γ concentrations (IU/ml) of the four participants with discordant QFT-GIT results
| Subject | Nu-4850 #1 | Nu-4850 #2 | PortaTherm™ |
|---|---|---|---|
| 1 Nil | 0.08 | 0.06 | 0.1 |
| Mitogen | 19.05 | 16.03 | 20.31 |
| Antigen | 1.14 | 0.83 | 0.25 |
| Antigen-nil | 1.06 | 0.77 | 0.15 |
| Interpretation | Positive | Positive | Negative |
| 2 Nil | 0.6 | 0.72 | 0.23 |
| Mitogen | 21.28 | 21.34 | 21.78 |
| Antigen | 1.19 | 1.38 | 0.56 |
| Antigen-nil | 0.59 | 0.66 | 0.33 |
| Interpretation | Positive | Positive | Negative |
| 3 Nil | 0.1 | 0.12 | 0.11 |
| Mitogen | 5.62 | 9.15 | 7.59 |
| Antigen | 0.3 | 0.52 | 0.29 |
| Antigen-nil | 0.2 | 0.4 | 0.18 |
| Interpretation | Negative | Positive | Negative |
| 4 Nil | 0.11 | 0.17 | 0.11 |
| Mitogen | 13.83 | 14.39 | 16.08 |
| Antigen | 0.63 | 0.14 | 0.11 |
| Antigen-nil | 0.52 | −0.03 | 0.00 |
| Interpretation | Positive | Negative | Negative |
IFN-γ = interferon-gamma; IU = international unit; QFT-GIT = QuantiFERON®-TB Gold In-Tube.
Analysis of absolute values
Quantitative measurements of IFN-γ levels for the three tubes (nil, mitogen and TB antigens) are determined from absolute assay OD values and reported as international units of concentration (IU/ml). As the absolute values of IFN-γ concentrations are available for all three tubes, these data were also compared between the three incubator trials. There were no significant differences for nil, mitogen or TB antigen tube IFN-γ concentrations between the two conventional NU-4850 incubators (P = 0.77, P = 0.28, P = 0.22, respectively) nor for the NU-4850 #1/PortaTherm or NU-4850 #2/PortaTherm comparisons (P = 0.24, P = 0.40, P = 0.88, and P = 0.14, P = 0.96, P = 0.56, respectively; raw data are graphically presented in Figures 4 and 5).
Figure 4.
Correlative plot of IFN-γ concentrations (TB antigen minus Nil values, IU/ml) for repeated QFT-GIT assays incubated in the same conventional incubator (NU-4850); R2 = 0.91438. IU = international unit; IFN-γ = interferon-gamma; QFT-GIT = QuantiFERON®-TB Gold In-Tube.
Figure 5.
Individual QFT-GIT results reported as IFN-γ concentrations (IU/ml) for three repeat QFT-GIT assays. PortaTherm values are represented as filled circles and repeated assays performed in the NU-4850 incubator are presented as light and dark triangles. Comparison of IFN-γ concentrations (IU/ml) between the three incubator trials is performed along the vertical axis. Any apparent clustering or linear regression of the data is an artefact of data arrangement and should not be considered in drawing comparisons of data. TB = tuberculosis; IU = international units; QFT-GIT = QuantiFERON®-TB Gold In-Tube; IFN-γ = interferon-gamma.
DISCUSSION
There are three significant findings from our comparison. First, the study demonstrates that the PortaTherm maintains an appropriate internal temperature (between 37° and 38°C) during the 20-h incubation period of QFT-GIT blood samples; indeed, greater temperature fluctuation was observed in the conventional NU-4850 incubator, probably attributable to the opening and closing of the incubator that occurs with general laboratory use. Second, there was a low frequency of disagreement between the positive/negative results for the three parallel tests (92% agreement). Discordance occurred in four of the 50 subjects and was due to small variations in IFN-γ response, as determined from absolute OD values. Test agreement between the three QFT-GIT assays proved to be higher than previously reported in a similar South African study in which three QFT-GIT assays were performed within 3 days of each other (two assays performed in parallel on Day 1 and the third assay on Day 3), yielding discordant results in five of the 27 participants.8 Finally, statistical analysis of the absolute IFN-γ concentrations of the three QFT-GIT assays of all 50 subjects determined no significant difference in the mean quantitative results between the three trials (NU-4850 #1/NU-4850 #2, NU-4850 #1/PortaTherm and NU-4850 #2/PortaTherm). We therefore determine that the PortaTherm incubator can be used to reliably incubate whole blood samples used in QFT-GIT testing without affecting the results or performance of the assay.
The strengths of this study include the use of repeat trials of the commercially available IGRA, the QFT-GIT assay, using identical and standard blood collection and laboratory techniques recommended by the manufacturer, thus making results more relevant to practicing clinicians. In addition, the blood samples used in performing repeat trials of the QFT-GIT assay were collected at the same time by the same technician. The three parallel assays were performed simultaneously in the same laboratory by the same technician, thereby eliminating inter-laboratory and inter-technician variability between trials. Finally, quantitative measurements of IFN-γ concentrations and positive/negative diagnoses for the three QFT-GIT assays of all 50 subjects were available at the time of testing, with no indeterminate results.
The sample population included occupational risk groups in a high TB burden setting, and is thus an appropriate group for study; the sample size permitted detection of large differences in incubator performance, and we cannot exclude the possibility that minor differences might have been missed; however, as the temperature stability of the PortaTherm is clearly demonstrated, we do not believe that this is likely. Finally, the PortaTherm represents an experimental or ‘prototype’ portable incubator, whose commercial use has not yet been established.
The 4% of subjects with QFT-GIT+/QFT-GIT− discordant results between the two repeated assays incubated in the same conventional incubator (NU-4850 #1 and NU-4850 #2) raises concerns about assay variability and the overall reproducibility of the QFT-GIT assay. How should such results be reported? In a recent study of 294 HIV-infected individuals, 20% of the study’s positive QFT-GIT results were reclassified as negative with repeated testing of the initial sample (intra-assay variability), and 32% of subjects previously diagnosed as positive received at least one negative test result after duplicate testing (inter-assay variability).9 Discordance of duplicate QFT-GIT assays warrants further investigation of assay variability, including inter-assay variability in negative and indeterminate results. Previous comparisons of the results from different generations of QFT tests have also raised questions regarding assay variability. Mahomed et al. found only fair agreement (75%) between QFT-GIT and QuantiFERON-TB Gold (QFT-G) results. Although the QFT-GIT assay and the earlier generation, QFT-G, rely on different methods of blood collection, both assays measure the response to virtually identical TB antigens and are therefore expected to yield similar test results.9,10 Based on the data presented, we expect repeated QFT-GIT assays to yield discordant dichotomous outcomes in 4% of tested patients. The apparent inter-assay variability inherent to QFT-GIT testing raises concerns about assay reproducibility and how IFN-γ responses are interpreted in making patient diagnoses. This is an additional important factor to be taken into consideration when evaluating, interpreting and trying to understand the within-patient temporal changes seen with serial QFT-GIT testing.8,11 Further investigation of inter- and intra-assay variability in QFT-GIT testing is needed to better understand the reproducibility and limitations of the test.
The results of this study also have implications for further investigational studies and clinical applications using the PortaTherm electricity-free incubator. The PortaTherm should be examined for its potential in performing similar diagnostic tests requiring incubation at 37°C in remote settings, as is being currently undertaken for use in performing the microscopic-observation drug susceptibility assay for rapid liquid culture-based diagnosis of TB and multidrug-resistant TB. Continued investigations in design and standardization of the PortaTherm are also of merit for future implementation and exploitation of this novel, low-cost technology. It would also be desirable to demonstrate the robustness of performance in the field to evaluate whether moving the PortaTherm around has an adverse effect on maintenance of temperature equilibrium. Heat leakage around the lid is minimal provided the box/lid seal is tight, and as phase-change material transitions in response to temperatures below 37°C it turns milky, providing an early indication that the packages require reheating.
In summary, the variability between results of QFT-GIT assays performed in the PortaTherm and conventional incubator was no greater than that between the two assays performed in the conventional incubator, indicating that the PortaTherm is a suitable tool for incubating QFT-GIT whole-blood samples in remote settings where access to a laboratory or electricity is limited. Further research into the reproducibility and inter- and intra-assay variability of the QFT-GIT assay is needed, and in particular, given the margin of error, to explore whether it might be more appropriate to define results within a defined distance either side of the cut-off as borderline rather than positive/negative. To estimate test variability, repeated assays should be performed on the same samples and on different samples from the same subject. We also recommend further investigation into other clinical uses of the PortaTherm as well as standardization of the design and standard operating procedures.
Acknowledgements
The authors wish to thank all participants; S Bird, F Rogoz, T Guo, J Rojas and S Fitzwater for important work on earlier iterations of the PortaTherm, M Zimic for statistical advice, J Coronel for laboratory coordination, E Ticona and M Ñavincopa for field coordination at Hospital Nacional Dos de Mayo; P Maguina for administrative support and D Juanito. MD was supported by a J William Fulbright US Student Scholarship administered by the Institute of International Education (IIE) and the United States Department of State. DAJM is supported by The Wellcome Trust (078067/Z/05/Z) and is also grateful for support from the National Institute for Health Research Biomedical Research Centre funding scheme and mentor support from B Gilman and J Friedland.
PortaTherm is a non-commercialized invention of AS.
References
- 1.Pai M, Zwerling A, Menzies D. Systematic review: T-cell-based assays for the diagnosis of latent tuberculosis infection: an update. Ann Intern Med. 2008;149:177–184. doi: 10.7326/0003-4819-149-3-200808050-00241. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Menzies D, Pai M, Comstock G. Meta-analysis: new tests for the diagnosis of latent tuberculosis infection: areas of uncertainty and recommendations for research. Ann Intern Med. 2007;146:340–354. doi: 10.7326/0003-4819-146-5-200703060-00006. [DOI] [PubMed] [Google Scholar]
- 3.Pai M, Joshi R, Dogra S, et al. Serial testing of health care workers for tuberculosis using interferon-gamma assay. Am J Respir Crit Care Med. 2006;174:349–355. doi: 10.1164/rccm.200604-472OC. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Perry S, Sanchez L, Yang S, Agarwal Z, Hurst P, Parsonnet J. Reproducibility of QuantiFERON-TB Gold In-Tube assay. Clin Vaccine Immunol. 2008;15:425–432. doi: 10.1128/CVI.00398-07. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Carrara S, Vincenti D, Petrosillo N, Amicosante M, Girardi E, Goletti D. Use of a T cell-based assay for monitoring efficacy of antituberculosis therapy. Clin Infect Dis. 2004;38:754–756. doi: 10.1086/381754. [DOI] [PubMed] [Google Scholar]
- 6.Avrami M. Kinetics of phase change. I. General theory. J Chem Phys. 1939;7:1103–1112. [Google Scholar]
- 7.Farid MM, Khudhair AM, Razack SAK, Al-Hallaj S. A review on phase change energy storage: materials and applications. Energy Conv Manage. 2004;45:1597–1615. [Google Scholar]
- 8.Detjen AK, Loebenberg L, Grewal HM, et al. Short-term reproducibility of a commercial interferon-gamma release assay. Clin Vaccine Immunol. 2009;16:1170–1175. doi: 10.1128/CVI.00168-09. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Luetkemeyer AF, Charlebois ED, Flores LL, et al. Comparison of an interferon-gamma release assay with tuberculin skin testing in HIV-infected individuals. Am J Respir Crit Care Med. 2007;175:737–742. doi: 10.1164/rccm.200608-1088OC. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Mahomed H, Hughes EJ, Hawkridge T, et al. Comparison of mantoux skin test with three generations of a whole blood IFN-gamma assay for tuberculosis infection. Int J Tuberc Lung Dis. 2006;10:310–316. [PubMed] [Google Scholar]
- 11.van Zyl-Smit RN, Zwerling A, Dheda K, Pai M. Withins ubject variability of interferon-γ assay results for tuberculosis and boosting effect of tuberculin skin testing: a systematic review. PLoS One. 2009;4:e8517. doi: 10.1371/journal.pone.0008517. [DOI] [PMC free article] [PubMed] [Google Scholar]