. 2011 May 12;2(6):1524–1538. doi: 10.1364/BOE.2.001524

Table 1. Differences in corneal layer boundary segmentation between two manual graders for 40 B-scans (Column I), as compared to the position differences between the automatic segmentation and the expert manual grader of the same 40 B-scans (Column II). Each pixel is approximately 3.4 µm in the cornea. The automatic segmentation has the same or lower mean difference and standard deviation compared to the expert than another trained manual grader.

 
Comparison of Two Manual Graders
Comparison of Automatic
and Manual Expert Segmentation
 
Column I
Column II
Corneal
Layer Boundary Mean Difference
(pixels) Standard Deviation
(pixels) Mean Difference
(pixels) Standard Deviation
(pixels)
Epithelium-Air Interface 1.5 0.4 0.6 0.4
Epithelium-Bowman Interface 1.2 0.5 0.9 0.5
Endothelium-Aqueous
Interface 2.1 2.3 1.7 2.1