Skip to main content
. Author manuscript; available in PMC: 2011 Jun 15.
Published in final edited form as: Am J Vet Res. 2011 Feb;72(2):203–209. doi: 10.2460/ajvr.72.2.203

Table 2.

Accuracy of CARD, CHROM, GEL, and SLIDE methods of feline blood typing as compared with the TUBE gold-standard method.

Typing method (%) No. of cats tested A antigen detection B antigen detection Overall accuracy (%)

Sensitivity (%) Specificity (%) Sensitivity (%) Specificity (%)
CARD 58 93.2 100.0 95.7 97.1 91.4
CHROM 58 97.7 100.0 95.7 97.1 94.8
GEL 490 100.0 100.0 100.0 99.3 99.4
SLIDE 487* 100.0 100.0 98.8 99.0 99.0
*

Data for the SLIDE test were calculated after exclusion of 3 of the 490 blood samples that could not be determined through this method because of persistent autoagglutination.