Table 3.

Multi-imager compatibility checklist

Criteria Test Desired Achieved Passed/Failed
Lab Test Motion Accuracy < 0.1 ° 0.024° Passed
Size Volume < 0.35 L 0.09 L Passed
Magnetism (GE Magnet Test (20)) Lifting force < 125 N 0.0 N Passed
Safe to test with ultrasound imager Passed
Clinical Engineering Safe to test in X-ray Fluoroscopy Imagers Clinical Engineering Sticker Passed
Safe to test in CT Scanners Passed
Safe to test in MRI Scanners Passed
Ultrasound Passive EP < 2% 0.145800% Passed
Active EA < 1% 0.148240% Passed
X-ray Passive EP < 2% 0.022400% Passed
Active EA < 1% 0.088200% Passed
Image Distortion Tests CT Scanner Passive EP < 2% 0.001500% Passed
Active EA < 1% 0.000032% Passed
MRI Passive EP < 2% 0.000188% Passed
Active EA < 1% 0.000187% Passed
Open MRI Passive EP < 2% 0.083000% Passed
Active EA < 1% 0.072000% Passed
Multi-Imager Compatibility Passed