Table 3.
Criteria | Test | Desired | Achieved | Passed/Failed | |
---|---|---|---|---|---|
Lab Test | Motion Accuracy | < 0.1 ° | 0.024° | Passed | |
Size | Volume < 0.35 L | 0.09 L | Passed | ||
Magnetism (GE Magnet Test (20)) | Lifting force < 125 N | 0.0 N | Passed | ||
Safe to test with ultrasound imager | Passed | ||||
Clinical Engineering | Safe to test in X-ray Fluoroscopy Imagers | Clinical Engineering Sticker | Passed | ||
Safe to test in CT Scanners | Passed | ||||
Safe to test in MRI Scanners | Passed | ||||
Ultrasound | Passive | EP < 2% | 0.145800% | Passed | |
Active | EA < 1% | 0.148240% | Passed | ||
X-ray | Passive | EP < 2% | 0.022400% | Passed | |
Active | EA < 1% | 0.088200% | Passed | ||
Image Distortion Tests | CT Scanner | Passive | EP < 2% | 0.001500% | Passed |
Active | EA < 1% | 0.000032% | Passed | ||
MRI | Passive | EP < 2% | 0.000188% | Passed | |
Active | EA < 1% | 0.000187% | Passed | ||
Open MRI | Passive | EP < 2% | 0.083000% | Passed | |
Active | EA < 1% | 0.072000% | Passed | ||
Multi-Imager Compatibility | Passed |