Table 2.

The Fill Volumes of Albuterol for a Jet Nebulizer and the I-neb Adaptive Aerosol Delivery (AAD) System Required to Achieve Similar Predicted Lung Doses

Jet nebulizer fill volume 2.0 mL 2.5 mL 3.0 mL 4.0 mL 5.0 mL
Jet nebulizer output to filter 429 μL 732 μL 892 μL 1370 μL 1836 μL
Predicted lung dose of albuterol from jet nebulizer
404 μg
688 μg
838 μg
1288 μg
1726 μg
Equivalent I-neb AAD System fill volume
0.42 mL
0.65 mL
0.77 mL
1.12 mL
1.47 mL
I-neb AAD System output to filter 333 μL 577 μL 691 μL 1034 μL 1359 μL
Predicted lung dose of albuterol from I-neb AAD System 420 μg 728 μg 870 μg 1302 μg 1712 μg