Quality control results for each analytical batch.

Series no. Blank
average
ng ml−1
Blank
S.D.,
ng ml−1
Solution
detection
limit
Blank
spike
recovery,
%
Sample
recovery,
%
NIST
Oyster
SRM, μgg−1
NRCC
Dolt-2
SRM,
μgg−1
NRCC
Dorm-2
SRM,
μgg−1
X-check
standard,
1 ngml−1
X-check
standard,
2 ngml−1
1 0.046 0.003 0.010 95 108 0.066 a 0.98 2.67
2 0.040 0.001 0.002 87 105 0.072 1.93 4.84 0.92 2.37
3 0.028 0.012 0.037 100 86 0.068 2.08 4.44 0.95 2.42
4 0.129 0.023 0.070 100 80 0.054 2.08 4.77 0.89 2.48
5 0.133 0.004 0.013 92 88 0.043 2.03 4.57 1.00 2.43
6 0.072 0.000 0.000 97 85 0.054 2.05 4.62 0.98 2.37
7 0.072 0.010 0.030 93 88 0.052 2.23 3.24 1.02 2.32
8 0.073 0.003 0.010 106 155 0.052 1.95 4.53 0.95 2.49
9 −0.019 0.035 0.104 103 91 0.095 2.55 5.15 1.05
10 −0.035 0.067 0.202 104 94 0.117 2.29 4.76 0.97
11 −0.009 0.007 0.020 101 102 0.076 2.60 5.19 0.97
Average Certified (SRMs) 0.048 0.015 0.045 98 98 0.068 2.18 4.61 0.97 2.44
0.064 ±0.007 1.99±0.10 4.64 ±0.26