. 2011 May;18(5):776–782. doi: 10.1128/CVI.00547-10

Table 2.

Specificity was tested by spiking QC2 into 6 negative serum batches and comparing the result to that for nontreated QC2

Sample Titer (EU/ml) Accuracy (%) Bias (%)
QC2 278
QC2 + serum 1 279 100.3 0.3
QC2 + serum 2 278 99.7 −0.3
QC2 + serum 3 269 96.6 −3.4
QC2 + serum 4 279 100.2 0.2
QC2 + serum 5 272 97.9 −2.1
QC2 + Ad35-positive serum 283 101.5 1.5