Skip to main content
. 2011 May;18(5):776–782. doi: 10.1128/CVI.00547-10

Table 4.

Reproducibility of Crucell and ABL combineda

Parameter IC1 IC2 IC3 IC4 IC5
Nominal titer (EU/ml) 635 282 125 55.8 24.8
Acceptance titer range (EU/ml) 476–794 226–339 100–151 44.6–66.9 18.6–31.0
No. of samples (Crucell + ABL) 29 29 30 30 48
Mean titer (EU/ml) 662.9 278.4 127.6 56.1 25.5
Reproducibility
    SD 86.4 39.8 13.7 8.1 2.9
    % CV 13.0 14.3 10.8 14.4 11.5
a

IC1 to IC5, internal control samples 1 to 5, respectively.