Table 2.

Concordance of HPV detection with Merck multiplex HPV PCR and standard Linear Array following DNA extraction with the Qiagen Spin blood kit

Assay No. of specimens with multiplex PCR/standard Linear Array result of:
Agreement rate Ppos Pneg P valuea
+/+ +/− −/+ −/−
HPV6 13 19 0 347 0.950 0.578 0.973 <0.001
HPV11 3 7 0 369 0.982 0.462 0.991 <0.001
HPV16 28 26 1 324 0.929 0.675 0.960 0.023
HPV18 9 20 0 350 0.947 0.474 0.972 <0.001
HPV31 6 39 0 334 0.897 0.235 0.945 <0.001
HPV33 1 18 0 360 0.953 0.100 0.976 <0.001
HPV35 9 14 0 356 0.963 0.563 0.981 <0.001
HPV39 9 39 0 331 0.897 0.316 0.944 <0.001
HPV45 7 6 3 363 0.976 0.609 0.988 0.505
HPV51 21 33 1 324 0.910 0.553 0.950 <0.001
HPV52 17 25 0 337 0.934 0.576 0.964 <0.001
HPV56 3 40 0 336 0.894 0.130 0.944 <0.001
HPV58 13 16 0 350 0.958 0.619 0.978 <0.001
HPV59 16 11 1 351 0.968 0.727 0.983 0.009
Any of the 14 HPVs 118 145 1 115 0.615 0.618 0.612 <0.001
β-globin 379 49 0 4 0.887 0.939 0.140 <0.001
a

P value as determined by McNemar's. P values of <0.05 indicate significant differences (in boldface).