. 2011 May;49(5):1899–1906. doi: 10.1128/JCM.00235-10

Table 4.

Concordance of HPV detection with Merck multiplex HPV PCR and standard Linear Array following DNA extraction with the Qiagen MinElute media kit

Assay No. of specimens with multiplex PCR/standard Linear Array result of:
Agreement rate Ppos Pneg P valuea
+/+ +/− −/+ −/−
HPV6 24 3 0 140 0.982 0.941 0.989 0.248
HPV11 11 1 0 155 0.994 0.957 0.997 1.000
HPV16 44 3 1 119 0.976 0.957 0.983 0.617
HPV18 19 4 3 141 0.958 0.844 0.976 1.000
HPV31 18 4 2 143 0.964 0.857 0.979 0.683
HPV33 3 6 4 154 0.940 0.375 0.969 0.752
HPV35 2 1 9 155 0.940 0.286 0.969 0.027
HPV39 4 1 14 148 0.910 0.348 0.952 0.371
HPV45 10 3 4 150 0.958 0.741 0.977 1.000
HPV51 8 1 26 132 0.838 0.372 0.907 <0.001
HPV52 14 4 13 136 0.898 0.622 0.941 0.052
HPV56 3 1 12 151 0.922 0.316 0.959 0.006
HPV58 3 6 1 157 0.958 0.462 0.978 0.131
HPV59 0 0 18 149 0.892 0.000 0.943 <0.001
Any of the 14 HPVs 122 5 35 5 0.760 0.859 0.200 <0.001
β-globin 167 0 1 0 0.994 0.997 0.000 1.000
a

P value as determined by McNemar's test. P values of <0.05 indicate significant differences (in boldface).