Table 1.
Specimen | Reconstituted reagent stored at −70°C | Mean CT ± SEM following storage interval of: |
|||
---|---|---|---|---|---|
0 wkb | 1 wk | 3 wk | 5 wk | ||
Positive lysatec | Master mix | 30.51 ± 1.09 | 30.30 ± 1.00 | 30.29 ± 0.95 | 30.84 ± 1.14 |
Negative lysatec | Master mix | 36.29 ± 0.21 | 35.99 ± 0.21 | 35.94 ± 0.23 | 35.99 ± 0.13 |
None | Positive controld | 34.72 ± 0.08 | 34.55 ± 0.09 | 34.55 ± 0.09 | 34.43 ± 0.08e |
None | Negative controld | 36.04 ± 0.04 | 35.90 ± 0.05 | 35.84 ± 0.05f | 35.81 ± 0.07f |
Mean cycle threshold values generated by analysis of reconstituted BD GeneOhm Cdiff assay master mix and control material frozen for up to 5 weeks at −70°C. The cycle threshold (CT) values derived from negative lysates and negative-control material indicate internal control amplification.
Freshly reconstituted master mix.
Ten lysates were incubated with frozen master mix from each storage interval.
Twenty replicates each of control material frozen for 1 week, 3 weeks, and 5 weeks.
Significantly different (P = 0.02) from the value for freshly reconstituted positive control.
Significantly different (P ≤ 0.009) from the value for freshly reconstituted negative control.