Table 1.

Comparison of the Abbott RealTime HR HPV and the HR Hybrid Capture 2 findings stratified for controls and cases

Sample Assay result (no. of samples)
Total
Abbott RealTime HR HPV HR HPV HC2
+
Controls + 52c 18 70
16 828c 844a
    Total 68 846b 914
Cases (≥CIN2) + 81f 0 81d
1 2f 3
    Total 82e 2 84
a

Absolute specificity for the Abbott RealTime HR HPV test was 92.3% (95% CI, 90.4 to 94.0).

b

Absolute specificity for HR HC2 was 92.6% (95% CI, 90.7 to 94.2).

c

Overall agreement for controls was 96.3% with a k value of 0.73 (95% CI, 0.65 to 0.82).

d

Absolute sensitivity for the Abbott RealTime HR HPV test was 96.4% (95% CI, 89.9 to 99.3).

e

Absolute sensitivity for HR HC2 was 97.6% (95% CI, 91.7 to 99.7).

f

Overall agreement for cases was 96.4% with a k value of 0.79 (95% CI, 0.402 to 1.00).