Skip to main content
. 2011 Apr;49(4):1446–1451. doi: 10.1128/JCM.02311-10

Table 2.

Agreement and disagreement among 998 specimens between the Abbott RealTime HR HPV and the HR HC2 results by HC2 signal intensity ratio

HR HPV HC2 score (RLU/CO) No. of agreements between Abbott RealTime HR HPV and HR HC2 (%) No. of disagreements between Abbott RealTime HR HPV and HR HC2
HR HC2 negative/Abbott RealTime HR HPV positive (%) HR HC2 positive/Abbott RealTime HR HPV negative (%)
<0.20 467/467 (100) None
0.21–0.30 245/246 (99.6) 1/246 (0.4)
0.31–0.50 92/100 (92.0) 8/100 (8.0)
0.51–0.99 26/35 (74.3) 9/35 (25.7)
1–5 23/31 (74.2) 8/31 (25.8)
>5 110/119 (92.4) 9/119 (7.6)