Table 3.
Results of reproducibility between Abbott RealTime HR HPV first and second tests among 521 samplesb
Second test | First test (no. of samples) |
Total (no. of samples) | |
---|---|---|---|
Positive | Negative | ||
Positive | 234 (44.9)a | 3 (0.6) | 237 (45.5) |
Negative | 5 (1.0) | 279 (53.5)a | 284 (54.5) |
Total | 239 (45.9) | 282 (54.1) | 521 (100) |
The agreement between the first and second test was 98.46%, with an overall k value of 0.97 (CI 95%, 0.95 to 0.99).