. 2011 Apr;49(4):1313–1317. doi: 10.1128/JCM.02555-10

Table 1.

Comparison of six treponemal assays to the FTA assay using serum specimensa

Assay and result FTA result (no. of samples)
% Sensitivity (95% CI) % Specificity (95% CI) % Agreement (95% CI) κ value
Positive Negative
BioPlex syphilis IgGb
    Positive 94 3 96.9 (90.9, 99.3) 98.5 (95.6, 99.7) 98.0 (95.6, 99.2) 0.96
    Negative 3 202
TP-PA
    Positive 93 5 95.9 (89.5, 98.7) 97.6 (94.3, 99.1) 97.0 (94.4, 98.5) 0.93
    Negative 4 201
Trep-Chek IgG
    Positive 93 1 95.9 (89.5, 98.7) 98.5 (95.6, 99.7) 97.7 (95.2, 99.0) 0.95
    Negative 4 203
    Equivocal 0 2
Trep-Sure
    Positive 94 4 96.9 (90.9, 99.3) 94.7 (90.6, 97.1) 95.4 (92.3, 97.3) 0.90
    Negative 3 195
    Equivocal 0 7
Trep-ID
    Positive 94 2 96.9 (90.9, 99.3) 99.0 (96.3, 100) 98.4 (96.1, 99.4) 0.96
    Negative 3 204
ViraBlot IgG
    Positive 91 2 93.8 (86.9, 97.4) 98.5 (95.6, 99.7) 97.0 (94.4, 98.5) 0.93
    Negative 5 203
    Equivocal 1 1
a

A total of 303 samples were analyzed.

b

Only 302 samples were analyzed for the BioPlex, as one sample did not yield a result due to an instrument error code.