Table 1.

The Optiscanner Fulfills ISO15197 Standard in Hypo- and Hyperglycemia. Distribution of Difference, Absolute, and Relative Absolute Difference within Each of Three Categories of Glucose Levels [Interquartile Range (IQR)]. ISO Accuracy Was Met in 96.7% for Hypoglycemia (<75 mg/dl), and in 99.2% and 99.5% for Normo- and Hyperglycemia (>180 mg/dl), Respectively

Hypoglycemic (<75 mg/dl glucose) n = 306 Difference (mg/dl) Median -7

Mean -5.4

Interquartile range [-10, -3]

Range [-16, 46]

Absolute difference (mg/dl) Median 8

Mean 7.7

IQR [4, 10]

Range [0 46]

Relative absolute difference (%) Median 12.5

Mean 13.2

IQR [7.6, 17.5]

Range [0.00, 63.01]

Euglycemic (≥75–≥180 mg/dl glucose) n = 480 Difference (mg/dl) Median -7

Mean -5.4

IQR [-10, -2]

Range [-25, 44]

Absolute difference (mg/dl) Median 7

Mean 7.8

IQR [4, 11]

Range [0, 44]

Relative absolute difference (%) Median 6.31

Mean 6.8

IQR [3.30, 9.40]

Range [0, 31.40]

Hyperglycemic (>180 mg/dl glucose) n = 369 Difference (mg/dl) Median -6

Mean -6

IQR [-11, -2]

Range [-119, 51]

Absolute difference (mg/dl) Median 8

Mean 9.4

IQR [4, 11]

Range [0, 119]

Relative absolute difference (%) Median 3.14

Mean 3.6

IQR [1.63, 4.88]

Range [0, 25.27]