Figure 3.
Example of pump limited compensation for uncertainty. At time “Now” the controller injects insulin and is then told at “Future measurement” whether the actual value was higher, at, or lower than the expected value (black Xs). The higher trajectory is rejected quickly by injecting more insulin. The lower value is corrected slowly by endogenous glucose production.