Skip to main content
. Author manuscript; available in PMC: 2012 Oct 1.
Published in final edited form as: Magn Reson Med. 2011 Mar 28;66(4):1089–1099. doi: 10.1002/mrm.22908

Figure 7.

Figure 7

Relative homogenous SNR as a function of matrix size for flexTPI acquisition with TTOTAL = 4 (a) and 10 (b) minutes, respectively. SNR values are relative to the homogenous SNR of tissue for TTOTAL=4 min, NM=44 and FR=0.603. Results for GMAX=GMAXSlew and both the symmetric and equatorial patterns of k-space trajectories at two different slew rates (90 mT/m/ms, 150 mT/m/ms) are shown.